Applications
- Medicine storage rooms: under Rules No. 260n humidity must not exceed 65 % unless the manufacturer states otherwise — Smartag RHT records both temperature and humidity.
- Temperature and humidity mapping when the protocol sets humidity criteria.
- Pharmacies and storerooms instead of a separate thermometer and hygrometer.
- Transport of humidity-sensitive cargo.
Specifications
| Temperature measurement range | −30 to +70 °C |
| Maximum permissible error, temperature | ±0.5 °C from −30 to +40 °C inclusive; ±0.7 °C above +40 to +70 °C |
| Relative humidity measurement range | 5 to 95 % |
| Maximum permissible error, humidity | ±5 % |
| Sensor | built-in digital, temperature and humidity |
| Memory | 45,000 values |
| Logging interval | configurable |
| Use | reusable |
| Indication | three LEDs, Start/Stop button |
| Connection | USB |
| Verification interval | 1 year |
| Verification procedure | MP-779/01-2024 |
| Operating conditions | −30…+70 °C, 84–106.7 kPa, humidity 5–100 % non-condensing |
| Dimensions and weight | max. 89 × 52 × 40 mm, max. 55 g |
| Average service life | at least 5 years |
| User software | Smartag User, in Russian and English |
| Instrument type | Smartag recording temperature indicators, State Register No. 91745-24 |
| Manufacturer | Hladverk Group LLC |
How long the memory lasts
| Logging interval | 45,000 values last for |
|---|---|
| 1 minute | about 31 days |
| 5 minutes | about 156 days |
| 10 minutes | about 312 days |
| 15 minutes | about 15 months |
How to use
- Set up
Connect the device to a computer and set the logging interval, thresholds and delayed start in the software.
- Start
Press Start or wait for the delayed start — the LEDs confirm that logging is on.
- Logging
The device records values at the set interval; the LEDs show whether a threshold was exceeded.
- Report
Connect to USB and save the report and data; record the serial number in the log.
- Next cycle
Clear the memory and start again — the device is reusable. Keep track of the verification date.
Modifications with humidity have a 1-year verification interval (Smartag T has 2 years). Plan verification in advance so the device does not drop out of a mapping study.
Frequently asked questions
Is Smartag RHT reusable?
Yes. After a cycle the data is saved, the memory is cleared and the device is started again.
What is the humidity accuracy?
±5 % in the 5–95 % range. Temperature: ±0.5 °C from −30 to +40 °C.
How often is Smartag RHT verified?
Every year, like all Smartag modifications that measure humidity.