1. Home
  2. Smartag Viewer

Smartag Viewer — processing data from Smartag devices

Smartag Viewer collects data from all devices and turns it into material you can attach to a report: charts, statistics for each point, summary tables and ready-made protocols. The software is designed for temperature mapping, so it includes what general-purpose viewers usually lack — mean kinetic temperature, grouping by zone and a protocol template.

Download the softwarefile coming soonUser guidefile coming soon
SystemWindows 10 or 11, 64-bit; administrator rights required for installation

What the software does

Chart for each pointtemperature and humidity on one scale, with limits and excursions marked
Statisticsminimum, maximum, mean, mean kinetic temperature, duration and number of excursions
Summary across several devicesall points of one zone on one screen, compared with each other
Grouping by zonepoints are grouped the same way as in the mapping protocol
Protocola ready Word or PDF document in the format accepted for mapping reports
Data identificationthe device serial number, recording period and settings are transferred to the report automatically

Mean kinetic temperature is not the same as the arithmetic mean: it gives more weight to time spent at elevated temperatures. When assessing the impact of an excursion on a product, it is this value, together with the duration of the excursion, that is considered.

How to use it

  1. Collect the files

    — connect the devices and copy the reports into one folder. With many devices, it is convenient to use a USB hub and copy everything in one go.

  2. Load the data

    — open the folder in the software. Devices are recognized by their serial numbers.

  3. Assign points to zones

    — exactly as they were placed during mapping: the logger layout must match the report.

  4. Check the data

    — for gaps, whether the period matches the study time, and whether any points were physically located somewhere other than recorded.

  5. Generate the report

    — the software compiles charts, tables and statistics into a document. Conclusions and justifications are written by a specialist, not by the software.

How this relates to the mapping methodology

The software covers data processing but does not replace the methodology. How the logger grid is designed, how long the study lasts and what counts as an acceptable result are explained on the pages “Warehouse mapping”, “How many data loggers you need” and “Re-mapping”. If you need round-the-clock control rather than a one-off report, see our monitoring system.

Other software and guides are in the software and guides section.

Frequently asked questions

Can I load data from devices made by other manufacturers?
The software is designed for Smartag devices. If you have a mixed fleet of devices, write to us: this is usually solved by data conversion, but it is discussed for the specific format.
Does the software decide by itself whether the study has passed?
No, and this is intentional. It calculates statistics and plots charts, while the conclusion on compliance is made by a specialist based on the acceptance criteria defined before the study started.
Can the report from the software be included in inspection materials?
The report is part of the package. It is accompanied by the study protocol, the logger layout map, copies of the devices' verification certificates and the conclusion. The full list is on the warehouse mapping page.

Need help with Smartag Viewer or software tailored to your task? Write to us

Request a quote

We will call you back

Pick a day and time (Moscow time) and our manager will call you.

When to call