| System | Windows 10 or 11, 64-bit; administrator rights required for installation |
What can be configured
| Logging interval | the period between measurements. For immunobiological medicinal products under SanPiN 3.3686-21, at least once per minute; for warehouse mapping, 1–5 minutes is usually sufficient |
| Alarm limits | the boundaries of the acceptable range; the device flags excursions and records their duration |
| Delayed start | a delay before recording begins, so that the period before the chamber is loaded and closed does not end up in the data |
| Identification | trip, zone or batch number — so the report can be linked to the right object |
| Connection check | the serial number is read from the connected device and shown on screen |
The serial number shown by the software is the same as the one on the device housing and in the verification certificate. You can use it to find the certificate on the serial number search page.
How to use it
- Install the software
— run the installer as administrator and follow the prompts. When you first connect a device, Windows may ask to install a driver.
- Connect the device
— with a cable or directly to a USB port. The software detects the device and shows its serial number and current settings.
- Set the parameters
— interval, limits, delayed start, identifier. Take the parameters from the mapping protocol or SOP rather than using the defaults.
- Write the settings to the device
— and make sure they appear when you read the device again.
- Check before going on site
— that the device has been started and the LED indicates recording mode. With a delayed start, recording begins later: this is normal.
Troubleshooting
| What happens | Likely cause | What to do |
|---|---|---|
| The software does not see the device | a charge-only cable without data lines | use a full USB data cable or connect the device directly |
| The device appears as a removable drive, but the software cannot find it | it is in use by another program or File Explorer | close any windows showing this drive and reconnect |
| The settings are not written | the device has already been started | stop recording, then configure it again |
| The driver does not install | no administrator rights | run the installer as administrator |
| Date and time in the report are shifted | the computer's time zone | check the time on the computer before configuring |
Other software and guides are in the software and guides section.